Beginning with Band 4 there is also a cold multiplier after the WCA, whose output power is specified in the ICD between the receiver cartridge and the First LO.

Band 1:

WCA: the output power of either channel shall be a minimum of 13 dBm. (20mW)

Band 2:

WCA:

Band 3:

WCA: the output power of either channel of the Band 3 first local oscillator shall be adjustable from –4 to +2 dBm in steps of 0.5 dB or less.

Band 4:

WCA: The output power of either channel shall be as follows: 15 mW over the 66.5 to 75 GHz range (adjustable in less than 0.25 dB steps down to 3.75 mW), 30 mW over 75 to 77.5 GHz range (adjustable in less than 0.25 dB steps down to 7.5 mW).

ICD: FEND-40.02.04.00-40.10.04.00-B-ICD: Output of each channel shall be adjustable from 200 μW to 20 μW in 0.5 dB steps. This requires that the loss in the waveguide between the WCA and the coolable frequency doubler input flange is as per the specifications that follow in this table. The cartridge manufacturer is responsible for supplying and installing this waveguide. It shall be possible to adjust the output of each channel from 20 μW down to 0 μW.

Band 5:

WCA: The Band 5 LO shall provide two LO power outputs (one for each mixer polarization) with a power level ≥ 0.4 mW.

ICD: FEND-40.02.05.00-40.10.05.00-C-ICD: Output of each channel shall be adjustable from 250 μW to 125 μW in steps not larger than 0.3 dB. This requires that the loss in the waveguide between the WCA and the coolable frequency doubler input flange is as per the specifications that follow in this table. The cartridge manufacturer is responsible for supplying and installing this waveguide. It shall be possible to adjust the output of each channel from 125 μW down to 0 μW.

Band 6:

WCA: The output power of either channel shall be adjustable from 5 mW to 20 mW in less than 0.5 dB steps.

ICD: FEND-40.02.06.00-40.10.06.00: 37.5 uW to 150 uW adjustable in 0.5 dB steps, provided that the waveguide loss between the WCA LO waveguide interface and the cryogenic tripler input waveguide is less than 2.0 dB (TBC).

Band 7:

WCA: the output power of either channel shall be adjustable from 1 mW to 3 mW in less than 1 dB steps, and from 3 mW to 8 mW in less than 0.5 dB steps.

ICD: FEND-40.02.07.00-40.10.07.00-B03-ICD: Output of each channel shall be adjustable from 150 μW to 37.5 μW in 0.5dB steps, from 37.5 μW to 20 μW in 0.8 dB steps, and from 20 μW to 10 μW in 1.0 dB steps. This requires that the loss in the waveguide between the WCA and the coolable frequency tripler input flange is as per the specifications that follow in this table. The cartridge manufacturer is responsible for supplying and installing this waveguide. It shall be possible to adjust the output of each channel from 10 μW down to 0 μW.

Band 8:

WCA: the output power of either channel shall be adjustable from 20 mW to 80 mW in less than 0.3 dB steps. In addition, from 65.5-70.0 GHz, the output power shall be adjustable up to 90 mW.

ICD: FEND-40.02.08.00-40.10.08.00-A-ICD: Output of each channel shall be adjustable from 20 μW to 80μW (goal shall be 100 μW) in 0.5 dB steps. This requires that the loss in the waveguide between the WCA and the coolable frequency quintupler input flange is as per the specifications that follow in this table. The cartridge manufacturer is responsible for supplying and installing this waveguide. It shall be possible to adjust the output of each channel from 20 μW down to 0 μW.

Band 9:

WCA: the output power of either channel shall be adjustable from 50 mW to 100 mW in less than 0.3 dB steps.

ICD: FEND-40.02.09.00-40.10.09.00-A08-ICD: Output of each channel shall be adjustable from 10 μW to 40 μW in 0.5dB steps. This requires that the loss in the waveguide between the WCA and the coolable frequency nonupler input flange is as per the specifications that follow in this table. The cartridge manufacturer is responsible for supplying and installing this waveguide. It shall be possible to adjust the output of each channel from 10 μW down to 0 μW.

Band 10:

WCA: the output power of either channel shall be adjustable from 20 mW to 60 mW in less than 0.5 dB steps for the frequency range of 88.3 GHz to <98 GHz, and from 20mW to 80 mW in less than 0.5 dB steps from 98 GHz to 104.7 GHz.

ICD: FEND-40.02.10.00-40.10.10.00-A-ICD: Output of each channel shall be adjustable from 15 μW to 3.75μW in 0.5 dB steps. This requires that the loss in the waveguide between the WCA and the coolable frequency nonupler input flange is as per the specifications that follow in this table. The cartridge manufacturer is responsible for supplying and installing this waveguide. It shall be possible to adjust the output of each channel independently from 3.75 μW down to 0 μW. (If the output power level is larger than 15 μW, then the step size will be 0.5 dB or finer.)

-- ToddHunter - 2010-02-01
Topic attachments
I Attachment Action Size Date Who Comment
2006-7-28ALMA-90.00.00.00-001-A-SPE.pdfpdf 2006-7-28ALMA-90.00.00.00-001-A-SPE.pdf manage 366 K 2017-12-27 - 18:43 ToddHunter ALMA Scientific requirements
2008-01-31-optical-analysis-final-report.pdfpdf 2008-01-31-optical-analysis-final-report.pdf manage 11 MB 2014-03-20 - 09:13 ToddHunter Full TICRA report
2008-08-20-ALMA-55.07.00.00-70.35.30.00-C-ICD.pdfpdf 2008-08-20-ALMA-55.07.00.00-70.35.30.00-C-ICD.pdf manage 292 K 2021-11-05 - 16:05 ToddHunter ICD between Back-End/FTS and Computing/Control
2012-09-03_ALMA-80.04.00.00-0038-A-SPE.pdfpdf 2012-09-03_ALMA-80.04.00.00-0038-A-SPE.pdf manage 345 K 2021-06-01 - 12:03 ToddHunter Sramek and Napier memo
2018-01-25-ALMA-60.00.00.00-70.40.00.00-E-ICD.pdfpdf 2018-01-25-ALMA-60.00.00.00-70.40.00.00-E-ICD.pdf manage 3 MB 2021-06-28 - 00:07 ToddHunter ICD from Correlator to Computing IPT
99142O.pdfpdf 99142O.pdf manage 5 MB 2020-09-24 - 23:17 ToddHunter  
ALMA-62.00.00.00-001-A-SPE.pdfpdf ALMA-62.00.00.00-001-A-SPE.pdf manage 539 K 2021-11-16 - 14:17 ToddHunter ACA specs
Band3_detail_FEND-40.01.03.00-011-A-REP.pdfpdf Band3_detail_FEND-40.01.03.00-011-A-REP.pdf manage 954 K 2014-03-27 - 10:11 ToddHunter 2009-03-06
Band3_optimize_FEND-40.01.03.00-012-A-REP.pdfpdf Band3_optimize_FEND-40.01.03.00-012-A-REP.pdf manage 606 K 2014-03-27 - 10:11 ToddHunter 2009-03-06
Band_10_Specifications_PY_KS_101027.pdfpdf Band_10_Specifications_PY_KS_101027.pdf manage 288 K 2011-01-31 - 13:55 ToddHunter Band 10 specs (Oct 2010)
FEND-40.02.01.00-0007-A02-SPE Technical Specifications.pdfpdf FEND-40.02.01.00-0007-A02-SPE Technical Specifications.pdf manage 530 K 2020-07-06 - 20:08 ToddHunter Band 1 technical specifications
FEND-40.02.04.00-062-A-CRE.pdfpdf FEND-40.02.04.00-062-A-CRE.pdf manage 213 K 2010-03-09 - 01:13 ToddHunter  
FEND-40.02.06.00-060-I-DSN.pdfpdf FEND-40.02.06.00-060-I-DSN.pdf manage 5 MB 2021-10-20 - 16:12 ToddHunter Band 6 overall design document
FEND-40.02.09.00-002-B-SPE_released.pdfpdf FEND-40.02.09.00-002-B-SPE_released.pdf manage 462 K 2009-11-11 - 13:49 ToddHunter FEND-40.02.09.00-002-B-SPE
IntelStratix.pdfpdf IntelStratix.pdf manage 1 MB 2021-09-09 - 15:17 ToddHunter comparison of TX and SX family
STR-C Released ALMA-80.04.00.00-005-C-SPE-2.pdfpdf STR-C Released ALMA-80.04.00.00-005-C-SPE-2.pdf manage 2 MB 2021-05-24 - 10:44 ToddHunter system level technical requirements
SYSE-80.04.01.00-004-R-DWG-04-01-2021.pdfpdf SYSE-80.04.01.00-004-R-DWG-04-01-2021.pdf manage 1 MB 2021-09-08 - 14:59 ToddHunter ALMA block diagram
Screen_Shot_2014-05-14_at_10.11.44_PM.pngpng Screen_Shot_2014-05-14_at_10.11.44_PM.png manage 401 K 2014-05-15 - 00:14 ToddHunter ALMA ACD photo mounted in antenna but with cryostat not present
acd_spie.pdfpdf acd_spie.pdf manage 342 K 2014-05-14 - 23:51 ToddHunter  
alma_calibration_device.pptxpptx alma_calibration_device.pptx manage 1 MB 2011-11-30 - 13:04 ToddHunter diagrams of ACD
cryo.pdfpdf cryo.pdf manage 75 K 2011-04-25 - 18:07 ToddHunter page 16 of Carter's optics design report showing layout of receiver windows
stratixModelNumber.pngpng stratixModelNumber.png manage 289 K 2021-09-09 - 14:42 ToddHunter  
systemDescription2004feb20.pdfpdf systemDescription2004feb20.pdf manage 291 K 2016-05-11 - 15:00 ToddHunter  
timeSynchronizationDesignRules.pdfpdf timeSynchronizationDesignRules.pdf manage 20 K 2016-05-11 - 15:00 ToddHunter  
version1.pdfpdf version1.pdf manage 253 K 2021-11-16 - 14:13 ToddHunter BLC specifications
Topic revision: r126 - 2021-11-16, ToddHunter
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding NRAO Public Wiki? Send feedback