Jay's student noticed that Tsys was periodically very high in their data and so they reported the problem. From looking at their data, I can confirm the following:
The data are frequency switched observations taken in VEGAS observing mode 23
The high Tsys values always fall in the first integration of each scan (both polarizations)
Resolution
For now, recommended that they blank the first integration of each scan