VEGAS dimple

The dimple

When testing the new fix to the Tsys Variations, it was noted that in position switched data a dimple exists in the center of strong continuum sources, as shown by the sharp drop in antenna temperature just above 8.58 GHz in the figure below. Note that position switched data use two scans, one on source and one off source, which are calibrated using the formula ([on/off-1]*Tsys - where Tsys is a scalar).

w3zoom4dimple.png

This dimple is fixed in channel space, i.e. it is located at the center of the bandpass regardless of LO setting or VEGAS mode (figure below - raw tp scan, bin number vs. counts).

rawdimplewithspike.png

Upon inspection of the individual scans, it becomes obvious that the negative dimple in position switched data is actually caused by a bump in the individual scans which does not appropriately calibrate out. To understand why it does not calibrate out, Ron Maddalena looked at the height of the dimple vs total power around the dimple, and found a surprisingly nice linear relation, as shown in the plot below.

image002.png

This non-zero y-intercept in the fit of the line is likely the cause of the dimple in the position switched data. The dimple does not appear to be resolution dependent. Below shows the dimple in two total power scans in modes 7 and 9 (L1/LBW1 100 MHz bandwidth, 32768 and 131072 channels, Fs=2e8). The baseline has been fit and removed around the center of the band, leaving a decent approximation of the dimple. The data have also been smoothed with a 100 bin boxcar. The height and width seems reasonably comparable between the two modes.

tpbump mode7vs9.png

When using a different bandwidth (and thus different sampling frequency), the scale of the dimple changes drastically. Below shows the same two dimples with the dimple from mode 6 overplotted (L1/LBW1 187.5 MHz bandwidth, 131072 channels, Fs=3.75e8). The mode 6 dimple is much larger.

tpbump modes6v7v9.png

One of our initial thoughts was that this could be a windowing function acting on the center Fs/4 spike. As an intial check of this, I looked at the spike height vs input power, though this trend goes in the opposite direction as the dimple height: The spike intensity decreases with total power in the spectral window (each color is a different scan, the individual dots represent individual integrations within the scan - note the logarithmic x axis):

pwrVSheight.png

This was a troubling relation for me. If I understand the Fs/4 spike correctly, it is the clock running the four ADCs leaking into the data. This to me would suggest that the spike height should be constant. I would also think that the spike height and the center bump would follow similar trends, but not the possibly linear and logarithmic trends shown above. What could be the cause of this bump? Could it be a drift in the clock driving the ADCs (though would that create the DC offset seen in the dimple height?)?

-- AdamKobelski - 2015-05-22

Topic attachments
I Attachment Action Size Date Who Comment
image002.pngpng image002.png manage 9 K 2015-05-22 - 14:14 AdamKobelski Total Power vs Dimple height
pwrVSheight.pngpng pwrVSheight.png manage 40 K 2015-05-22 - 14:28 AdamKobelski  
rawdimplewithspike.pngpng rawdimplewithspike.png manage 27 K 2015-05-22 - 19:42 AdamKobelski  
tpbump_mode7vs9.pngpng tpbump_mode7vs9.png manage 41 K 2015-05-22 - 14:19 AdamKobelski  
tpbump_modes6v7v9.pngpng tpbump_modes6v7v9.png manage 40 K 2015-05-22 - 14:24 AdamKobelski  
w3zoom4dimple.pngpng w3zoom4dimple.png manage 548 K 2015-05-22 - 13:28 AdamKobelski  
This topic: CICADA > CICADAGreenBankSpectrometerDimple
Topic revision: 2015-05-22, AdamKobelski
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding NRAO Public Wiki? Send feedback