Test behavior of Stokes U and V data with a new BOF file that uses registers in place of delay elements.
Details
Session begins at 10:00
Switched VEGAS to pulsar branch
Will use 00-CoherentModeArtifactTesting in c1500x0512 mode (this is the mode the BOF file was built for). Will set-up at S-Band to try to avoid FFT overflow issues.
Taking one scan with current BOF file (c1500x0512_x14_7_2016_Feb_24_1437.bof) for comparison
SB submitted at 10:06
Scan is #1
SB ended at 10:08
Accidentally had vegas.fold_dumptime set to 1.0 seconds, so there are lots of sub-integrations. Data should be fine but going to take another scan anyway
Taking second scan with current BOF file, 10-s dump times.
SB submitted at 10:10
Scan is #2
SB ended at 10:11
Data don't look great. It's almost like there are drop-outs in the phase domain. Dropped packets? Something wrong in the PFB? Definitely seeing phase wraps in U/V, though.
Switching to test BOF file (c1500x0512_x14_7_rtd_2017_Mar_21_1121.bof)
SB submitted at 10:22
Scan is #3
First attempt failed to start HPC subprocess and aborted. Maybe a systemstop/start is needed.
Resubmitted at 10:26
Scan is #4
SB ended at 10:28
No obvious changes in U/V data
Going to take a scan using test BOF file but at L-Band. Will try changing fftshift to all f's and set vegas.scale to 110
SB submitted at 10:37
Scan is #5
SB ended at 10:39
We are getting dropped packets. Possible source of the features in the phase-domain?
Phase wrap in Stokes U and V data still seems to be present with new BOF file, but dropped packets could complicate interpretation.