TGBT14A_912_141 Check Modes 4-6 BOF file again

Goals

  • Test version 137 of L1/LBW1 BOF - l1_ver137_1500_511_fir_2015_Dec_10_1529.bof

Details

  • Start setting up at 10:00am
  • Start with X-band
  • Run 8bank_simple_Xband_mode4
  • Scan 1: hpc3 taking too long to be ready; scan abort
  • Scan 2: Looks good. Tsys around 85k, but weather is crap, lbw_gain = 2895887,2914360, 2914403, 2903478, 3061542, 3228283, 2991856, 3117166
  • Both polns, and all IFs look good. Total powers (counts) change by a factor of 2, but that should be ok. Tsys = ~ 77 +/- 0.1 and 84+/- 0.5. This should be ok.
  • Scan 3 is some unknown thing that somehow happened!

  • L-band mode 4,5,6,7
  • Scan 4: Mode 4
  • Scan 5: Mode 5
  • Scan 6: Mode 6
  • Scan 7: Mode 7
  • All look good, and identical. Call it a good!
  • 10:50am - finito