Configure the system as normal for VEGAS and the Spectrometer, and look at the respective backend inputs with a Spectrum Analyzer. Look for anything which looks anomalous between the two.
Take a cable (or cables) which normally connect the AFR to the Spectrometer, and either tee them off or connect them into the inputs of the VEGAS IF modules. Take tp with cal data (i.e. cal diode switching) either simultaneously, or consecutively with the Spectrometer and VEGAS. Compare the results.
Take data with VEGAS with the IF system configured for Mode 1, but trick VEGAS into running with Mode 4 (i.e. same IF configuration all the way to the VEGAS ADCs). Compare Mode 1 and Mode 4 results.
As above, but try different settings for the IF rack bandpass filter, from all-pass down to 20 MHz. Look at the converter rack attenuator settings to confirm Steve's point about bandwidth and variance.
Repeat above looking at the noise source in the IF rack
Physically swap some Converter Modules (not just the cables). See whether the results move around with the converter modules or not.
Repeat the tests were we drive some VEGAS IF modules with identical copies of a single converter module through a splitter, and others with their normal converter rack inputs. Compare the results.
(suggestion from Ray) - use a narrow band mode, and spread it across the wide band IF.
Details
Set up at 8:00am
Ray needs to swap versions, Galen sets up Spectrum Analyzer
Weather is terrible; thick grey cloud, rain...
Spectrum Analyzer Checks
Run a tp spectrum to make sure system is configured correctly, and leave exactly this way for visual inspection.
VEGAS Mode 1
bank_gain_mode1_xband - Accidentally had A, not E - need to redo E.
archivist
balance - this does not seem to balance the IF rack! - fiddle around manually, with cleo, then run balance again
SetValues (5.644)
Scan 1: Track - but set up wrong. Decide we want to use the same converter modules for Spectrometer and VEGAS - A,B,C,D
bank_gain_mode1_xband_acs - confirms we will be using A,B,C,D
bank_gain_mode1_xband - hi cals running! Proceed with this anyway....
Galen is seeing 10dB "spikes" appearing in the spectrum analyzer.
Switch inputs in IF Rack to be wideband noise source. Rebalance, but do not otherwise configure
Galen deduces the spikes are probably RFI.
ACS "Mode 1" (800 MHz)
bank_gain_mode1_xband_acs
balanceACS
Scan 3: Track: into Spectrometer. IF power levels 8,3,1 off by 0.85,0.54,0.51dB
Tsys (from Spectrometer): [66,64,67,65,66,64,67,65] - but did this with lo cal diode
Spectrometer Samplers 3 and 6 are faulty at the moment. Tee-off samples 1,2,7,8 to VEGAS.
Green path, green numbers
Connect up Converter Rack A CM1&5 and CM4&8 to Spectrometer J1, J2, J7, J8, Tee off to VEGAS A1,A2,B1,B2 (not convinced which order), straight into ADC
Make Spectrometer switching signal master for everything
Scan 4: Track. Need to fill spectrometer data manually
compare scan 4 to scan 2; scan four Tsys spectrum lines up much more closely than scan 2 between different IFs
Scan 5 - Track on IF wideband noise source (need to adjust converter rack attenuators) Nonsense; no noise source.
Scan 6 - simulates the "on". Turn switching signals off - still looking at noise source
Increase all attenuators by 1dB. After this, values:
CM1 1dB
CM5 5dB
CM4 9dB
CM8 6.5dB
Scan 7 - simulates the "off"
Check Scan 4
bank_gain_mode1_xband_acs, tp, low noise.
archivist
balanceACS - does not correctly balance Spectrometer, do it manually again.
CM1,5,4,8: 0,8,9.9.5 - sets levels good for VEGAS, slightly off for Spectrometer (1.24,0.98,0.84,0.77)
Scan 8 - Track
adjust CM attenuators so that Spectrometer is as close as realistic to 0.848 - this seems to be as close as we can get
VEGAS Tsys values 52.30,51.82,52.53,51.18
Spectrometer Tsys values 82.37, 85.5, 79.0, 75.1
red path, red numbers
Take the outputs of the AFR, and feed them into the inputs of the VEGAS IF Module. Need to take pads off input to Spectrometer, and add them to input to VEGAS IF modules
CM1,5,4,8 - 13.0,22.5,23,24dB
Scan 9 - Scan 9 seems significantly worse than Scan 8; Scan 8 in VEGAS agrees reasonably with Scan 4
Need to do this more systematically.
A,B, IFR connected to input of VEGAS IF module
IF Rack 17,20,16,19
CM rack 13.0, 22.5, 23, 24
VEGAS: -20.6,-20.6, -19,-19 (slowly varying)
Spec Duty cycles 0.83, 0.81,0.79,0.87
Scans 10-14 five "identical" scans simply submit Track 5 times
balanceACS - just does a "balance"
Sets IF rack to 16,20,16,19
CM rack 13.875, 22.625, 24.625, 23.875
VEGAS ~ -20.5, -20.5, -18. -19
Spec Duty cycles 0.84,0.88,0.97,0.82
Scans 15-19 five "identical" scans
From GBTIDL; Tcal p0 = 2.00, Tcal p1 = 1.88
Green path, green numbers again
A,B, IFR connected directly to input of ADC (switch pads around again).
balanceACS - Converter Rack fails
IF rack 14, 17, 14, 17
CM Rack 0,9,10,10
VEGAS -19,-20,-21,-20
Spec duty cycles 1.1,1.1,1.0,0.93
Scans 20-24 five "identical" scans
Red path, purple numbers
revert to path where AFR goes to input of VEGAS IF modules
On A, move 15dB pad from before ADC to before VEGAS IF module
Removed 15 dB from before Spectrometer
Remove 3dB pad from Bank A, 6dB pad from Bank B
balanceACS
IF rack 13, 16, 13, 16
CM 15,24,25,25
VEGAS -19.5,-19.5, -21,-20
Spec duty cycles 1.3,1.29,1.16,1.04
Scan 25-29 five "identical" scans
Put Everything back to normal
Configure just for Spectrometer: bank_gain_mode4_xband
Scan 30 - pol 1 in all Banks falls off and then starts recovering during the middle of a scan!! Mode 4 by mistake