VEGAS Testing 2014 August 26

Goals

  • Test the new mode 2/3 BOFs (h16k_ver112e_1500_64.bof and h16k_ver112e_1080_2014_Aug_21_1618.bof)
  • Look for spikes within mode 2/3 data, and look for their prevalence in other modes (to see if it is an postIF issue)
  • See if the "High Tsys in first integration" problem has been resolved

Details

  • Jason wired a noise source into the top of the rack, putting into the VEGAS IF system
  • Had VEGAS read the new BOF files
  • Created Mode2BankABankB
  • VEGAS fault, it did not like the BOF file for bank A, though it was also being used in Bank B (Bank B appeared to load acceptably)
    • restarting VEGAS did not solve it.
  • TGBT14A_912_21 - Scan 1: ifnum=0 is the noise source ifnum=1 is the standard IF system.
    • BOF files still FUBAR.
  • Scan 2: Noise source is out of the bandwidth
  • Many small spikes noted
  • Scan 3: for off-off
  • Scan 4 and 5: 600s scans in mode 3

This topic: CICADA > WebHome > GreenBankSpectrometer > CICADAGreenBankSpectrometerCommissioningLogs > CICADAGreenBankSpectrometer2014Aug26
Topic revision: 2015-01-27, RichardPrestage
This site is powered by FoswikiCopyright © by the contributing authors. All material on this collaboration platform is the property of the contributing authors.
Ideas, requests, problems regarding NRAO Public Wiki? Send feedback