Pol 0 History: | 2008-12-11 | Removed Mixer s/n 113 for gain slope issues; replaced with s/n 120; |
2009-03-05 | Removed Mixer s/n 120 for oscillation, replaced with Mixer s/n 127 | |
Pol 1 History: | 2008-12-11 | Removed Mixer s/n 104 for gain slope issues; replaced with s/n 138 |
2009-05-05 | Removed Mixer s/n 138; high noise temp. and low image rejection on second cool down, replaced with Mixer s/n 155 | |
2009-05-13 | Mixer Block 155 was rested? |
Notes |
Pol 0 |
Pol 1 |
IR's Pol 0/1 from CTS | ||
IR's from MTS | ||
Placeholder for Mixer block 155, test 2 | ||
NT's from CTS | ||
NT's from MTS | ||
Placeholder for Mixer block 155, test 2 | ||